Installationfor media temperature up to 400°C
Flexiblecapillary between rigid stem and housing
Electricalbuilt-in calibration
Various highlevel outputs
4-20 mA,2-wire 0-10 V DC, 3/4-wire and other volt output signal ranges
PerformanceCharacteristics Pressure Range: 0-17 bar to 0-2000 bar
Accuracy:MDT422F ±0.5% FSO (up to 50 bar ±1% FSO);
Repeatability:MDT422F ±0.1% FSO (up to 50 bar ±0.2% FSO); MDT462F ±0.2% FSO
Resolution:Infinite Maximum Overload (without in\"uencing operating data): 2x pressurerange for range 1000 and 1400
BurstPressure: 6x pressure range max. 3000 bar Material in Contact with Media: 15-5PH SST (Mat. No. 1.4545)
DyMax® coatedElectrical Characteristics Confguration: 4-arm
Wheatstonebridge strain gage Internal Shunt Calibration: 80% of full scale ±1% OutputSignal: 2-wire 4-20 mA
SupplyVoltage: 12-36 Vdc Zero Balance: ±1%
RezeroInitiation: Locally via zero potentiometer, or remotely via short circuitbetween pins “D” and “F” Range Calibration 80% FSO: “Short circuit” between connectionspins “E” and “F” at the sensor or externally from the control room
LoadResistance: Maximum 1200 at 36 Vdc Maximum 500 at 24 Vdc
IsolationResistance: 1000M* at 50 Vdc
TemperatureInfluence Diaphragm: Max.
Temperature:400ºC Zero Shift (due to temperature change): MDT422F X < 0.2 bar/10ºCMDT462F X < 0.4 bar/10ºC Housing: Max.
Temperature:85ºC Zero Shift (due to temperature change): ±0.2% FSO/10ºC
SensitivityShift (due to temperature change): MDT422 X ±0.1% f.s.v./10ºC up to 50 bar±0.2% f.s.v./10ºC MDT462 X ±0.4% f.s.v./10ºC
Approvals&Self Certifications CE: Directive 2004/108/EC ISO
Datasheet(s) [PDF]
- dynisco-mdt462f-1-2-35c-15-46-t80-sil2--298555.pdf