1. Products
  2. HEIDENHAIN
  3. HEIDENHAIN LIP 382 Exposed Linear Encoder

HEIDENHAIN

HEIDENHAIN LIP 382 Exposed Linear Encoder

Ask for Stock
WhatsApp

Contact with us! 24 business hours (excluding weekends and holidays) to confirm stock status and estimated delivery time.

Specifications

LIP 382

Measuring standard

Coefficient of linear expansion

DIADUR phase grating on Zerodur glass ceramic; grating period 0.512 µm
?therm = (0±0.1) x 10–6 K–1

Accuracy grade

±0.5 µm (higher accuracy grades upon request)

Baseline error

? ±0.075 µm/5 mm

Measuring length ML* in mm

70 120 150 170 220 270

Reference marks

No

Interface

  1 VPP

Integrated interpolation

Signal period

0.128 µm

Cutoff frequency –3 dB

? 1 MHz

Scanning frequency*

Edge separation a

Traversing speed

? 7.6 m/min

Interpolation error

Position noise RMS

±0.01 nm

0.06 nm (1 MHz1))

Laser

Scanning head and scale mounted: Class 1

Scanning head not mounted: Class 3B

Laser diode used: Class 3B

Electrical connection

0.5 m cable to interface electronics (APE), separate adapter cable (1 m/3 m/6 m/9 m)
connectable to APE

Cable length

See interface description, but ? 30 m (with HEIDENHAIN cable)

Power supply

DC 5 V ±0.25 V

Current consumption

< 190 mA

Vibration 55 Hz to 2000 Hz

Shock 11 ms

? 4 m/s2 (EN 60068-2-6)

? 50 m/s2 (EN 60068-2-27)

Operating temperature

0 °C to 40 °C

Mass Scanning head

Interface electronics

Linear scale

Connecting cable

150 g

100 g

ML 70 mm: 260 g, ML ? 150 mm: 700 g

38 g/m

* Please select when ordering

1) –3 dB cutoff frequency of the subsequent electronics


Name Download
heidenhain-lip-382-expose… (2).pdf Download